TEST PATIENT (Internal Incubation Mode)
Option 1 on the Main Menu screen allows the user to test a patient
sample. Follow the analyzer prompts.
The most recent USER ID is always displayed. Press ENTER to accept
the ID, or enter a new User ID. This field will accept 15 alpha or numeric
characters. To leave this field blank, delete the information using the ←
key.
The CASSETTE LOT# must be entered to proceed to the next step. For
convenience, the last 2 numbers only can be entered if the lot has not
changed. The CASSETTE LOT# is located on the cassette pouch. The
software requires that the lot number is entered in the correct format: one
alpha character followed by four numeric characters (e.g., L1002).
TLi
System User Manual
IQ
Section 4 − Software Functions - Detailed Descriptions
fFN MAIN MENU

1-TEST PATIENT

2-SET CALIBRATION
3-DAILY QC
TEST PATIENT
USER ID
>JOHN SMITH
ENTER-CONFIRM
TEST PATIENT
CASSETTE LOT#
>L1002
ENTER-ACCEPT