ITEM
PROBLEM
5
Printer not on when
test was run.
6
Printer output in
unusual font.
7
Analyzer turned off
after calibration, or
power failure
occurred after
calibration.
8
Cassette cannot be
removed.
TLi
System User Manual
IQ
Section 7 – Troubleshooting
SOURCE
Printer
Printer
Power
Analyzer
SOLUTION
Turn printer on. Recall the test result on the analyzer
display screen. Press PRINT/ENTER on the analyzer to
print the result.
Turn analyzer and printer off and back on.
The calibration remains in memory. Reset calibration only
if prompted by analyzer.
Turn analyzer off and back on to reinitialize the system.
7 – 3