3 MECHANICAL INSTALLATION

3.1 Siting Requirements

3.1.1 Instruments – Fig. 3.1
Caution.
Mount in a location free from excessive vibration.
Mount away from harmful vapours and/or dripping fluids.
Information. It
is
instrument at eye level, allowing an unrestricted view of
the front panel displays and controls
Maximum Distance
164 ft. (50m) – cell
constant <0.05
A – Maximum Distance of Instrument to Cell
–20°C
Min.
B – Within Temperature Limits
IP66
C – Within Environmental Limits
Fig. 3.1 Siting Requirements – Instrument
4
preferable
to
mount
the
Conductivity
Cell
55°C
Max.
3.1.2 Conductivity Cells – Fig. 3.2
Caution. Ensure that the integral cable (where
applicable) does not hang against hot or abrasive objects
when the plug is connected to the bulkhead socket.
Note. Allow sufficient clearance for easy removal of
cell for cleaning – see Section 3.4 for overall dimensions
of cells.
Model 2278
Max. Temperature – 230°F (110 °C)
A – Maximum Temperature Limits
Model 2278
Max. Pressure – 150 p.s.i.
B – Maximum Pressure Limits
Model 2278
C – Acid/Alkali Concentration Limits
Fig. 3.2 Siting Requirements – Conductivity Cell
Acid – 6% Concentration
Alkali – 8% Concentration