...8 CALIBRATION
...8.3 Factory Settings Page
x x x x x
Temp Zero (100R)
x x x x x
Temp Span (150R)
– – – – –
Adjust RTX Zero
– – – – –
Adjust RTX Span
– – – – –
Adjust RTX Zero2
– – – – –
Adjust RTX Span2
0 0 0 0 0
Alter Fact. Code
Return to
5 0 0
top of page
microsiemens/cm
24
Calibrate Temperature Zero
Set the temperature simulator to 100Ω and allow the instrument display to stabilize.
Advance to next parameter.
Calibrate Temperature Span
Set the temperature simulator to 150Ω and allow the instrument display to stabilize.
Advance to next parameter.
Adjust Retransmission 1 Zero
Set the milliammeter reading to 4mA.
Note. The retransmission range selected in the
not affect the reading.
Advance to next parameter.
Adjust Retransmission 1 Span
Set the milliammeter reading to 20mA.
Note. The retransmission range selected in the
not affect the reading.
Advance to next parameter.
Adjust Retransmission 2 Zero (available only on 4623/800 and 4628/800 instruments)
Set the milliammeter reading to 4mA.
Note. The retransmission range selected in the
not affect the reading.
Advance to next parameter.
Adjust Retransmission 2 Span (available only on 4623/800 and 4628/800 instruments)
Set the milliammeter reading to 20mA.
Note. The retransmission range selected in the
not affect the reading.
Advance to next parameter.
Alter Factory Code
Set the factory settings access code to a value between 00000 and 19999.
Return to
Operating Page.
does
Set Up Retransmission Page
does
Set Up Retransmission Page
does
Set Up Retransmission Page
does
Set Up Retransmission Page