13 Annual Product
Inspection
(Therapist, Technician, Leckey Product
Advisor, Dealer)
Leckey recommend that each product should
be subject to a detailed inspection at least
once a year and every time the product is
reissued for use. This inspection should
be carried out by a technically competent
person who has been trained in the use of
the product and should include the following
checks as a minimum requirement.
1.Check that there is no lateral movement or
wobble in any of the moving parts. Ensure all
adjustment knobs and bolts are in place and
secure.
2.Check all ratchet handles, knobs, nuts,
bolts and plastic buckles are in place,
replacing any missing items. Paying
particular attention to the following items;
> Headsupport locking bolts.
> Backrest height and angle adjustments
bolts.
> Seat depth adjustment bolts.
> Leeway Pelvic Cradle / hip guide
attachment bolts.
> Footrest height and angle adjustment.
3.Check the chassis height adjustment
mechanism is working properly. If the
chassis is foot pedal operated ensure the
seat height doesn't change when the pedal
is released. Also, check that the locking
pin engages securely to prevent accidental
height adjustment of the chassis.
4.Adjust the seat to its maximum range of
tilt in space and ensure that the locking lever
locks the seat out securely at varying point in
this range.
5.Check that where the seat and chassis join
there is no visible wear and tear on the metal
components.
6. Lift the base to check each castor
individually. Make sure they are moving freely
and remove any dirt from the rubber wheels.
Check that the brakes lock the wheels
securely
7.Visually check the structure of the product
paying attention to weld points on the frame
ensuring there are no signs of fatigue or
cracking around the welds.
8.Leckey recommend that a written record is
maintained of all annual product inspections.
If in any doubt to the continued safe
use of your Leckey product or if any
parts should fail, please cease using
the product and contact our customer
service department or your local dealer
as soon as possible.
14 Re-issuing Leckey
Products
Most Leckey products are assessed and
ordered to meet the needs of an individual
user. Before reissuing a product we
recommend that the therapist prescribing
the product has carried out an equipment
compatibility check for the new user and has
ensured that the product being re-issued
contains no modifications or special
attachments.
A detailed technical inspection should be
carried out on the product prior to re-issuing.
This should be carried out by a technically
competent person who has been trained in
the use and inspection of the product. Please
refer to section 11 for the required checks to
be carried out.
Ensure the product has been cleaned
thoroughly in accordance with section 12 of
this manual.
Ensure a copy of the user manual is supplied
with the product. A copy can be downloaded
from our website www.leckey.com
Leckey recommend that a written record is
maintained of all product inspections carried
out during the reissue of the product.
If in any doubt to the continued safe
use of your Leckey product or if any
parts should fail, please cease using
the product and contact our customer
service department or your local dealer
as soon as possible.

Advertisement

Table of Contents
loading

Table of Contents