Chapter 5 Calibration; Figure 5.2 Cold Junction Calibration Setup - Brainchild C21 User Manual

Auto-tune fuzzy / pid process / temperature controller
Table of Contents

Advertisement

5520A
5520A
Calibrator
Calibrator
K-TC
Figure 5.2
Figure 5.2
Cold Junction Calibration Setup
Cold Junction Calibration Setup
The 5520A calibrator is configured as K type thermocouple output
The 5520A calibrator is configured as K type thermocouple output
with internal compensation. Send a 0.00 C signal to the unit under
with internal compensation. Send a 0.00 C signal to the unit under
calibration.
calibration.
The unit under calibration is powered in a still-air room with
temperature 25 3 C. Stay at least 20 minutes for warming
up. Perform step 1 stated above, then press scroll key until
the display shows
40.00.
Press scroll key for at least 5 seconds. The display will blink a
moment and a new value is obtained . Otherwise , if the
display didn't blink or if the obtained value is equal to -5.00 or
40.00, then the calibration fails.
Perform step 7 to calibrate
*
if required.
Step 7.
Step 7.
Setup the equipments same as step 6. The unit under
calibration is powered in a still-air room with temperature
A B
A B
3 C.
3 C.
calibrator source is set at 0.00 C with internal compensation
mode.
A B
Stay at least 20 minutes for warming up . The
UM0C911C
C21
C91
K+
9
10
K
Stay at least 20 minutes in still-
Stay at least 20 minutes in still-
air room
air room
room temperature 25
room temperature 25
L
L
. Press up/down key to obtain
gain
gain
of
cold junction
cold junction
B
4
5
A
A
L
L
3 C
3 C
compensation
50
50