AGFA CR 30-X Service Manual page 381

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS171
.06E
IMPORTANT:
For digitizers which still have the old side
plate with straight edge installed:
It is recommended to replace the side plate
by the "Side Plate Transport Unit (left hand
side)" preventively.
Spare part number see section 1.6.
Installation instructions are part of delivery
of the spare part.
(10) Remove screw at erasure unit.
(11) Disconnect two cables at erasure
unit.
(12) Take out erasure unit.
CAUTION:
IP jams possible if IP guide plate is deformed.
Do not bend IP guide plate at removal.
(13) Slide IP guide plate to the left and lift
it out.
(14) Check condition of the 4 belts in the
cassette unit.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
01-2009
Cassette side plate with sloped rear edge
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Maintenance
Figure 10:
(new design)
Figure 11
Figure 12
Chapter 9 / 15
Agfa Company Confidential