AGFA CR 30-X Service Manual page 115

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(3)
If the version on the control PC is higher*, select the respective PCB(s)
(node(s)) in the Digitizer to update.
*If versions on control PC and digitizer are identical, no software upload is required.
The situation "software on digitizer is higher than on control PC" may only happen during
installation due to logistical reasons. In this case it is mandatory to install the latest digitizer
service client on the control PC. Refer to section 2.1.
(4)
Select "Upload selected nodes".
(5)
Wait for approximately 7 minutes till message "versions are current" appears
in the message box.
(6)
Close the CR 30-X service client.
(7)
Answer the dialogue popping up "did you perform any repair or maintenance
activities" with "Yes".
(8)
Switch off the digitizer.
(9)
Restart the control PC.
(10) Switch on the digitizer.
(11) Wait till the digitizer is booted-up, i.e. till the status indicator is green.
(12) Perform a backup and store it on a safe place (not on the control PC!).
IMPORTANT:
The installed versions on the PMI board, IP Handling board (HDL Board) and Slow
Scan Board (SSC Board) have always to be the same.
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2010
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Tools and auxiliary Means
Agfa Company Confidential
Chapter 3.2 / 19