AGFA CR 30-X Service Manual page 225

Type 5175/100/105/110/120/125
Hide thumbs Also See for CR 30-X:
Table of Contents

Advertisement

DD+DIS150.06E
(10) Disconnect ground cable at right hand side.
(11) Remove front cover.
(12) Remove toothed belt.
ASSEMBLY:
(1)
Mount in reverse order.
CALIBRATION:
(1)
Perform a stall calibration for the opener motor.
Refer to chapter 3.6, Adjustments.
(1)
Insert cassette with unexposed IP in the digitizer.
Verification:
(2)
At the workstation select examination type "System diagnosis – Flatfield".
(3)
Select "ID" to identify the image: Scanning starts.
(4)
Confirm that the image arrives at the workstation.
(5)
Reject the image.
The Toothed Belt Stepper Motor Opener is replaced.
Result:
The controlled version of this document resides on MedNet. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2009
Replacements / Repair Procedures
DOCUMENT CONTROL NOTE:
CR 30-X
Type 5175 / 100/110
Repair and Service
Figure 206
Figure 207
Chapter 3.5 / 83