Dc Current Specifications - Fluke 5520A Manual

Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

Manual Supplement

DC Current Specifications

Range
0 to 329.999 μA
0 to 3.29999 mA
0 to 32.9999 mA
0 to 329.999 mA
0 to 1.09999 A
1.1 to 2.99999 A
0 to 10.9999 A
(20 A Range)
[1]
11 to 20.5 A
[1]
Duty Cycle: Currents <11 A may be provided continuously. For currents >11 A, see Figure B. The current may be provided 60-T-I
minutes any 60 minute period where T is the temperature in °C (room temperature is about 23 °C) and I is the output current in
Amps. For example, 17 A, at 23 °C could be provided for 60-23-17 = 20 minutes each hour. When the 5520A is outputting
currents between 5 and 11 amps for long periods, the internal self-heating reduces the duty cycle. Under those conditions, the
allowable "on" time indicated by the formula and Figure B is achieved only after the 5520A is outputting currents <5 A for the "off"
period first.
[2]
Specifications apply within two minutes of selecting operate.
Range
0 to 329.999 μA
0 to 3.29999 mA
0 to 32.9999 mA
0 to 329.999 mA
0 to 2.99999 A
0 to 20.5 A
Absolute Uncertainty,
tcal ±5 °C
±(ppm of output + μA)
90 days
1 year
120 + 0.02
150 + 0.02
80 + 0.05
100 + 0.05
80 + 0.25
100 + 0.25
80 + 2.5
100 + 2.5
200 + 40
160 + 40
380 + 40
300 + 40
380 + 500
500 + 500
[2]
800 + 750
1000 + 750
Bandwidth
0.1 Hz to 10 Hz
p-p
2 nA
20 nA
200 nA
2000 nA
20 μA
200 μA
Max Compliance
Resolution
Voltage
1 nA
0.01 μA
0.1 μA
1 μA
10 μA
10 μA
100 μA
[2]
100 μA
Noise
Bandwidth
10 Hz to 10 kHz
4
5520A Operator
Max Inductive
Load
V
mH
10
10
7
7
400
6
6
4
4
rms
20 nA
200 nA
2.0 μA
20 μA
1 mA
10 mA
7/11