Fluke 5520A Manual page 9

Hide thumbs Also See for 5520A:
Table of Contents

Advertisement

Manual Supplement
AC Voltage (Sine Wave) Specifications (cont.)
Range
Frequency
10 to 20 Hz
20 to 45 Hz
45 Hz to 1 kHz
10 to
329.999 mV
1 to 5 kHz
5 to 10 kHz
10 Hz to 30 kHz
10 to 20 Hz
20 to 45 Hz
45 Hz to 1 kHz
0.33 to
3.29999 V
1 to 5 kHz
5 to 10 kHz
10 to 30 kHz
10 to 20 Hz
20 to 45 Hz
45 Hz to 1 kHz
3.3 to 5 V
1 to 5 kHz
5 to 10 kHz
[1]
There are two channels of voltage output. The maximum frequency of the dual output is 30 kHz.
Note
Remote sensing is not provided. The AUX output resistance is < 1 Ω. The maximum load capacitance is 500 pF, subject to the
maximum burden current limits.
AUX (Auxiliary Output) [dual output mode only]
Absolute Uncertainty,
tcal ±5 °C
±(% of output + μV)
90 days
1 year
0.15 + 370
0.2 + 370
0.08 + 370
0.1 + 370
0.08 + 370
0.1 + 370
0.15 + 450
0.2 + 450
0.3 + 450
0.4 + 450
4.0 + 900
5.0 + 900
0.15 + 450
0.2 + 450
0.08 + 450
0.1 + 450
0.07 + 450
0.09 + 450
0.15 + 1400
0.2 + 1400
0.3 + 1400
0.4 + 1400
4.0 + 2800
5.0 + 2800
0.15 + 450
0.2 + 450
0.08 + 450
0.1 + 450
0.07 + 450
0.09 + 450
0.15 + 1400
0.2 + 1400
0.3 +1400
0.4 + 1400
[1]
Max
Resolution
Burden
±(% of output + floor)
1 μV
5 mA
10 μV
5 mA
100 μV
5 mA
8
5520A Operator
Max Distortion and
Noise
10 Hz to 100 kHz
Bandwidth
0.2 + 200 μV
0.06 + 200 μV
0.08 + 200 μV
0.3 + 200 μV
0.6 + 200 μV
1 + 200 μV
0.2 + 200 μV
0.06 + 200 μV
0.08 + 200 μV
0.3 + 200 μV
0.6 + 200 μV
1 + 200 μV
0.2 + 200 μV
0.06 + 200 μV
0.08 + 200 μV
0.3 + 200 μV
0.6 + 200 μV
7/11

Advertisement

Table of Contents
loading

Table of Contents