Set Fixed mA Output

If the operation mode is Communicator Only, this
function is not available and an error message is shown.
A warning to change the loop to MANUAL before you
proceed is shown. Push the selection knob to proceed.
Push  to go to the Trim, Set and Zero menu.
As the HART device is changed to fixed output mode, an
information screen is shown. An error is shown if the
HART device rejects the mode change command. When
the mode change is successful, the screen in Figure 38 is
shown.
Figure 38. Set Fixed mA Output Screen
This screen is used to set a fixed output and monitor the
result with the Product measurement. The measurement
is updated once a second.
The range of values that can be set is 3.0 mA to 21.0 mA.
1.
Use the selection knob to select and increment or
decrement the value in steps shown by the selected
decade.
2.
Push and hold the selection knob to send the new
value to the HART device and stay on this screen.
An error is shown if the HART device rejects the
value.
3.
Push  to change the HART device to normal
output mode and go to the Trim, Set, and Zero menu.
A warning to change the loop to AUTOMATIC is
shown first. An error is shown if the device rejects the
mode change command.
gzx47.eps
Precision Loop Calibrator
Trim, Set, and Zero Menu
37