Front / Remote / Serial Selection - Varian Turbo-V 300HT Instruction Manual

969-9424 969-9524
Hide thumbs Also See for Turbo-V 300HT:
Table of Contents

Advertisement

Press the CURRENT push-button and the display
shows:
I
= 0
.
0 0 A
X
.
.
K R P M
where:
I = is the DC current drawn by the pump
range (0.00 to 5.00 Ampere)
P= is the DC power drawn by the pump
(range 0 to 999 Watt)
KRPM = is the theoretical rotational speed of
the pump as a function of the controller output
frequency (range 3 to 42 KRPM)
°C = is the temperature of the outer ring of the
upper bearing (range 00 to 99°C)
X = during operation a selected set point
condition (1 or 2 contrast inverted) appears
when the programmed threshold speed value
is not reached.
Press the CYCLE NUMBER twice and the display
shows:
X X X X C Y C L E
P U M P
L I
P =
X X ° C
X X X X X m
F E
X X X X X h
where:
0 W

Front / Remote / Serial Selection

Press CYCLE NUMBER and PUMP CURRENT
pushbuttons together for at least 2 seconds and the
processor enters in a routine where it is possible to
program the controller.
In this routine, the CYCLE push-button is used for
choosing/changing the value or condition;
the PUMP CURRENT push-button is used to enter and
confirm the value.
At any time it is possible to exit this routine by pressing
the CYCLE and PUMP CURRENT pushbuttons at the
same time for at least 2 seconds.
The display shows:
C O N F I
S O F T W A R E
X X X
and then:
S O F T
S E L E C T I
where: XXX = YES or NO.
If YES is selected, the Soft Start mode allows the pump
to rump-up the Normal speed within ten steps.
When NO is selected, the Soft Start mode is
deselected and the rump-up of the pump will be done
within 300 seconds.
The controller is factory set to YES.
The Soft Start mode may be deselected/selected only
when the pump is stopped.
64
TECHNICAL INFORMATION
CYCLE = are the cycles performed (range 0 to
9999)
m = is the elapsed time related to the cycle
number displayed (range 0 to 99999 minutes)
PUMP LIFE = is the total operation time of the
pump (range 0 to 99999 hours).
G U R A T I
V E R S I
X X X X
S T A R T
O N :
NOTE
87-900-876-01 (B)
O N
O N
X X X X
M O D E
X X X

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Turbo-v 550

Table of Contents