Connection Procedure - Vega VEGABAR 14 Operating Instructions Manual

Pressure transmitter
Hide thumbs Also See for VEGABAR 14:
Table of Contents

Advertisement

Connect power supply
Connect via angle plug
connector acc. to DIN
43650
16

5.2 Connection procedure

Proceed as follows:
1 Loosen the screw on the rear of the plug connec-
tor
2 Remove plug connector and seal from VEGABAR
3 Lever the plug insert out of the plug housing
Cable entry
Plug insert
Fig. 6:
Loosen the plug insert
4 Dismantle the connection cable by approx. 5 cm,
strip the wire ends approx. 1 cm
5 Lead the cable through the cable entry and into
the plug housing
6 Connect the wire ends to the screw terminals acc.
to the wiring plan
Cable entry
Fig. 7:
Connection to the screw terminals
Plug housing
Plug housing
Plug insert
VEGABAR 14

Advertisement

Table of Contents
loading

Table of Contents