Data card status
messages
(continued)
Data card
operation
1.800.561.8187
Status Message
Fragmented
A fragmented FAT was detected. Monitoring cannot proceed with
a fragmented data card. Either change the data card or format to
continue.
Unformatted
Data card is not formatted. Format the data card or replace with a
different card to continue.
Invalid Card
Data card is not valid for monitoring. Change the data card to
proceed.
The data card screen prompts users to set filename and format card in preparation for
monitoring and writing of data.
Action...
STEP 1: From the Advanced
Options screen, press Next to
display data card status.
• If the default filename is
acceptable without change, and
if the card is formatted to save
data, press Finish. The sytem is
ready to begin monitoring. Go
to Section B Monitor Same
Circuit - Turning Monitoring
On/Off on page 5-24.
• Press Change to enter a new
site/file name. Proceed to Step
2 on page 5-22.
• Press Format card to format
the data card.
NOTE: All data and setups
stored in data card will be lost
when you format the card.
Copy any files that you want to
save to a computer first before
formatting card.
• Press Memory Card Contents
to view list of data files stored
in card.
• Press Cancel to discard
changes and return to Start
Menu.
www.
Description
Result...
information@itm.com
.com
CH 5/ Start Menu
MARK113pg
Continued on next page
5-21