OEM TUF-2000F User Manual page 25

Integrated ultrasonic flow meter

Advertisement

TUF-2000F User Manual
0221-0222
0229-0230
0231-0232
0233-0234
0257-0288
32
0289
0311
0313
1437
1438
1439
1440
1441
1442
1451
1521
1529
Note :(1) The internal accumulator is been presented by a LONG number for the integer part
together with a REAL number for the decimal fraction. In general uses, only the integer part
needs to be read. Reading the fraction can be omitted. The final accumulator result has a
relation with unit and multiplier. Assume N stands for the integer part (for the positive
accumulator, the integer part is the content of REG 0009, 0010, a 32-bits signed LONG
integer,), Nf stands for the decimal fraction part (for the positive accumulator, the fraction part
is the content of REG 0011, 0012, a 32-bits REAL float number,), n stands for the flow
multiplier (REG 1439).
then
The final positive flow rate=(N+Nf ) ×10 n-3 (in unit decided by REG 1438)。
The meaning of REG 1438 which has a range of 0~7 is as following:
2
Inner pipe diameter
2
Upstream delay
2
Downstream delay
2
Calculated travel time
LCD buffer
1
LCD buffer pointer
2
Worked time for today
2
Worked time for this month
1
Unit for flow rate
1
Unit for flow totalizer
1
Multiplier for totalizer
1
Multiplier for energy
accumulator
1
Unit for energy rate
1
Device address
2
User scale factor
2
Manufacturer scale factor
2
Electronic serial number
0 cubic meter
1 liter
2 American gallon
REAL4
REAL4
REAL4
REAL4
BCD
INTEGER
LONG
LONG
INTEGER
INTEGER
INTEGER
INTEGER
INTEGER
INTEGER
REAL4
REAL4
BCD
3
(m
)
(L)
(GAL)
25
0056
In millimeter
In microsecond
In microsecond
In microsecond
Unsigned, in
seconds
Unsigned, in
seconds
See note 5
Range 0~7,see
note 1
Range 0~7,see
note 1
Range 0~10,see
note 1
0=GJ 1=Kcal
2=KWh,3=BTU
Read only
High byte first