1. Introduction
1.1 Preface
Welcome to use new generation transit-time ultrasonic flow meter, please read the user manual carefully
before using.
1.2 Features
◆ Linearity: 0.5%, Repeatability: 0.2%, Accuracy:±1%
◆ Several type transducers for selection, measuring pipe size is from DN15mm to DN6000mm
Adopt low voltage, multi-pulse technology to improve accuracy, useful life and reliability.
Powerful Recording Function, record the totalizer data of the last 64 days/64 monthes/5 years.
1.3 Flow measurement principle
The TUF-2000F ultrasonic flow meter is designed to measure the fluid velocity of liquid within a closed
conduit. The transducers are a non-contacting, clamp-on type, which will provide benefits of non-fouling
operation and easy installation.
The TUF-2000F transit-time flow meter utilizes two transducers that function as both ultrasonic transmitters
and receivers. The transducers are clamped on the outside of a closed pipe at a specific distance from
each other. The transducers can be mounted in V-method where the sound transverses the pipe twice, or
W-method where the sound transverses the pipe four times, or in Z-method where the transducers are
mounted on opposite sides of the pipe and the sound crosses the pipe once. This selection of the mounting
method depends on pipe and liquid characteristics. The flow meter operates by alternately transmitting and
receiving a frequency modulated burst of sound energy between the two transducers and measuring the
transit time that it takes for sound to travel between the two transducers. The difference in the transit time
measured is directly and exactly related to the velocity of the liquid in the pipe, show as follows:
Downstream transducer
MD
T
flow
V
sin
2
T
T
Tdown
up
down
Tup
θ
Upstream transducer
spacing
Where
θ is the include angle to the flow direction
M is the travel times of the ultrasonic beam
1

Advertisement

loading

Summary of Contents for OEM TUF-2000F

  • Page 1 ◆ 1.3 Flow measurement principle The TUF-2000F ultrasonic flow meter is designed to measure the fluid velocity of liquid within a closed conduit. The transducers are a non-contacting, clamp-on type, which will provide benefits of non-fouling operation and easy installation.
  • Page 2 TUF-2000F User Manual D is the pipe diameter Tup is the time for the beam from upstream transducer to the downstream one Tdown is the time for the beam from downstream transducer to the upstream one ΔT=Tup –Tdown 1.4 Optional transducer ◆...
  • Page 3 TUF-2000F User Manual Analogue output: 4-20mA or 0-20mA current output. Impedance 0 1k. Accuracy 0.1%. Output OCT output: Frequency signal (1~9999HZ) Relay output: over 20 source signal (no signal, reverse flow etc.) RS485 serial port Three analogue input Input Three-wire PT100 resistor input (optional) Automatically record the totaliser data of the last 64 days / 64 months / 5 years;...
  • Page 4: Menu Window Details

    TUF-2000F User Manual 1. Menu Window Details The flow meter user interface comprises about 100 independent windows. That are numbered by M00, M01,M02……. M99,M+0,M+1 etc. You do not need to remember all the menu windows. Just remember the most command used windows and the proximate window numbers of some uncommonly used windows would be sufficient.
  • Page 5 TUF-2000F User Manual If the positive totalizer is turned off, the positive totalizer value shown on the screen is the value prior to its turn off Display flow rate and NEG(negative) totalizer If the negative totalizer is turned off, the negative totalizer value shown on the screen is the value prior to its turn off Display date and time, flow rate.
  • Page 6 TUF-2000F User Manual (3) Gasoline (4) Fuel oil (5) Crude Oil (6) Propane at -45C (7) Butane at 0C (8)Other liquids(need to enter sound speed in M21 and viscosity in M22) Diesel Oil (10)Caster Oil (11)Peanut Oil (12) #90 Gasoline (13) #93...
  • Page 7 TUF-2000F User Manual (2) Function to store the current parameters into the flash memory, so that these parameters will be solidified and will be loaded as the default parameters every time when power is turned on. Entry to store to or restore from the internal Flash memory, as many as 9 different...
  • Page 8 TUF-2000F User Manual display is used as the display device. Setup for local segmental LCD display. Enter 0 or 1 for the non-auto-scan mode; Enter 2~39 for the auto-scan mode. In the auto-scan mode the display will automatically scan displaying from 00 to the entered number of the local segmental LCD display.
  • Page 9 TUF-2000F User Manual (0) 4-20mA output mode (setup the output range from 4-20mA) (1) 0-20mA output mode (setup the output range from 4-20mA, This mode can only be used with Version-15 flow meter) (2) Serial port controls 0-20mA (3) 4-20mA corresponding fluid sound speed...
  • Page 10 TUF-2000F User Manual For Version-15 flow meter, you need to indicate on your orders that you need the frequency function; Otherwise you will get a flow meter which has no frequency output circuits. Window to setup the minimum flow rate value which corresponds to the lower frequency limit of the frequency output.
  • Page 11 TUF-2000F User Manual 20. Timed alarm #2 21.Batch Totalizer Full 22. Periodically M51 Timer 23. Oct Not Using The OCT circuit does not source voltage at its output. It must be connected with an external power and pull-up resistant for some occasions.
  • Page 12 TUF-2000F User Manual which stand for the total flow is over 100 cubic meters during the period of every day from 20:00 to 06:00, setups is like M51 start time =20:00:00 M51 interval =10:00:00 M51 log times =9999 (means always)
  • Page 13 TUF-2000F User Manual Displays the Time Ratio between the Measured Total Transit Time and the Calculated time. If the pipe parameters are entered correctly and the transducers are properly installed, the ratio value should be in the range of 100±3%. Otherwise the entered parameters and the transducer installation should be checked.
  • Page 14 TUF-2000F User Manual This function can used to produce an alarm or output when fluid material changes. Displays total flow for this month(only for the time past) Displays total flow for this year(only for the time past) Display the not-working total time in seconds. The total failure timer will also include the time when power off, if the back-up battery is applied.
  • Page 15 TUF-2000F User Manual (5) Liquid type (for non-standard liquids, the sound speed of the liquid is also needed) (6) Transducer type adapted to the flow meter. Generally the Standard M1 clamp-on transducers will be the selected option. (7) Transducer mounting methods (the V-method or Z-method is the common option) (8) Check up the Space displayed on M25 and install the transducers accordingly.
  • Page 16 TUF-2000F User Manual An optimum location would be defined as a straight pipe length full of liquid that is to be measured. The piping can be in vertical or horizontal position. The following table shows U p s t r e a m...
  • Page 17 TUF-2000F User Manual 4.3 Transducers Installation The transducers used by the TUF-2000 series ultrasonic flow meter are made of piezoelectric crystals both for transmitting and receiving ultrasonic signals through the wall of liquid piping system. The measurement is realized by measuring the traveling time difference of the ultrasonic signals. Since the difference is very small, the spacing and the alignment of the transducers are critical factors to the accuracy of the measurement and the performance of the system.
  • Page 18 TUF-2000F User Manual 4.3.4 Z-method Installation Z-method is commonly used when the pipe diameter is above 200mm. 4.3.5 W-method Installation W-method is usually used on plastic pipes with a diameter from 15mm to 50mm.
  • Page 19: Installation Checkup

    TUF-2000F User Manual 4.3.6 N-method Installation Rarely used method. 4.4 Installation Checkup Through the checkup of the installation, one can: check the receiving signal strength, the signal quality Q value, the traveling time difference of the signals, the estimated liquid speed, the measured traveling time of the signals and the calculated traveling time ratio.
  • Page 20: Communication Protocols

    TUF-2000F can also set to automatically output data at a period which is programmable. The TUF-2000F also has a special command sets to facilitate the use of the flow meter in a GSM network.
  • Page 21 By default, the RS232/RS485 will setup with 9600,none,8,1(9600bd,none parity,8 data bits, 1 stop bit) 5.1.1 MODBUS REGISTERS TABLE MODBUS REGISTERS TABLE for TUF-2000F (please take notice the difference with the water meter MODBUS table) REGISTER NUMBER...
  • Page 22 TUF-2000F User Manual 0019-0020 Positive energy decimal fraction REAL4 0021-0022 Negative energy accumulator LONG 0023-0024 Negative energy decimal REAL4 fraction 0025-0026 Net accumulator LONG 0027-0028 Net decimal fraction REAL4 0029-0030 Net energy accumulator LONG 0031-0032 Net energy decimal fraction REAL4...
  • Page 23 TUF-2000F User Manual 0062 Times for the beeper INTEGER Writable。Max 255 0062 Pulses left for OCT INTEGER Writable。Max 65535 0072 Error Code 16bits, see note 4 0077-0078 PT100 resistance of inlet REAL4 In unit Ohm 0079-0080 PT100 resistance of outlet...
  • Page 24 TUF-2000F User Manual 0115-0116 Positive accumulator REAL4 In Cubic Meter, float 0117-0118 Negative accumulator REAL4 In Cubic Meter, float 0119-0120 Net energy accumulator REAL4 In GJ,float 0121-0122 Positive energy accumulator REAL4 In GJ,float 0123-0124 Negative energy accumulator REAL4 In GJ,float...
  • Page 25 TUF-2000F User Manual 0056 0221-0222 Inner pipe diameter REAL4 In millimeter 0229-0230 Upstream delay REAL4 In microsecond 0231-0232 Downstream delay REAL4 In microsecond 0233-0234 Calculated travel time REAL4 In microsecond 0257-0288 LCD buffer 0289 LCD buffer pointer INTEGER 0311 Worked time for today...
  • Page 26 TUF-2000F User Manual 3 imperial gallon (IGL) 4 American million gallon (MGL) 5 Cubic feet (CF) 6 American oil barrel (1 barrel =42gallon) (OB) 7 Imperial oil barrel (IB) While The energy flow rate =(N+Nf )×10 n-4 (unit decided by REG 1441)...
  • Page 27 TUF-2000F User Manual Bit13 reserved Bit14 internal timer over flow Bit15 analog input over range Please try to override these energy-related bits first when in flow-only measurement, (5) Unit code for flow rate Cubic Cubic meter Cubic meter Cubic meter /day...
  • Page 28 TUF-2000F User Manual 0162 Data pointer Integer Range:0~63 2817 Day and Error Code Day in high byte 2818 Month and year Year in high byte 2819-2820 Total working time LONG Net total flow for the 2821-2822 REAL4 Net total energy for the...
  • Page 29 TUF-2000F User Manual 3339-3340 Total working time LONG 3341-3342 Net total flow for the REAL4 month 3343-3344 Net total energy for the REAL4 month 。。。。 。 。 。 。 。 。 。 。 。 。 。 。 。 。 。 。 。。。。。。。。。。。。。。 。。。。。。 。 。 。 。 。 。 。 。 。 。 。 。 。 。 。 。 。 。 。 。...
  • Page 30 3825-3840 5.2 The FUJI extended communication protocol TUF-2000F is compatible with the TUF7-FUJI extended communication protocol which used in our previous Version7 ultrasonic flow meters. This protocol is a set of basic commands that are in ASCII format, ending with a carriage return (CR) and line feed (LF), For most of the commands, The line feed (LF) should be better omitted for fast responding.
  • Page 31 TUF-2000F User Manual DIN(cr) Return net totalizer ±dddddddE±d(cr) DIE(cr) Return net thermal energy totalizer ±dddddddE±d(cr) DIE+(cr) Return positive energy totalizer ±dddddddE±d(cr) DIE-(cr) Return negtive energy totalizer ±dddddddE±d(cr) DIT(cr) Return net total flow for today ±dddddddE±d(cr) DIM(cr) Return net total flow for this month ±dddddddE±d(cr)
  • Page 32 TUF-2000F User Manual (cr) Output current ‘a’ mA at the AO output (cr)(lf) Note 6 terminal. BA1(cr) Return the resistance for T1 ±d.ddddddE±dd(cr)(lf) BA2(cr) Return the resistance for T2 ±d.ddddddE±dd(cr)(lf) BA3(cr) Return current value of AI3 (0~20mA) ±d.ddddddE±dd(cr)(lf) BA4(cr) Return current value of AI4 (0~20mA) ±d.ddddddE±dd(cr)(lf)
  • Page 33 TUF-2000F User Manual 2.d stand for digit 0~9, the number before ‘E’ is an integer. 3.Working status code, 1~6 letters, refer to error code related chapter. 4.‘s’ is ‘ON’,’OFF’ or ‘UD’ For example ‘TR:ON,RL:ON’ means the OCT and RELAY are closed ‘TR:UD,RL:UD”...
  • Page 34 TUF-2000F User Manual For example if the address number 88 flow meter is going to be addressed, the command should like: NXDV(CR), the decimal value of X should be 88. The prefix W is strongly recommended for new users. (3) The ‘W’ prefix...
  • Page 35 Flow meters made by Huizhong have more than 10 different communications protocols. For the easier replacement of a water meter, most of these protocols are realized in TUF-2000F flow meters. Here only one of them, the default for compatible protocols CRL-61D (D<=50mm), is given for reference.
  • Page 36 TUF-2000F User Manual (2) Reading the recorded meter data (command 49) Format:2Ah XXh 49h Answer: 26h XXh 49h LL(BCD 码) ZZh The difference between the command 4A and command 49 is that the late command reads out the data which are recorded in the meter by the time which is defined by command 4C.
  • Page 37 TUF-2000F User Manual 05h stand for there exist hardware failure, repair may needed. 5.4 Key Value Table The key values are used in a network application. By use of the key value and a command ‘M’, we can operate the flow meter through the network on a computer or other kind of terminals.