OEM TUF-2000F User Manual page 2

Integrated ultrasonic flow meter

Advertisement

TUF-2000F User Manual
D is the pipe diameter
Tup is the time for the beam from upstream transducer to the downstream one
Tdown is the time for the beam from downstream transducer to the upstream one
ΔT=Tup –Tdown
1.4 Optional transducer
-1 clamp-on type transducer, pipe size from DN15-100mm
TS
-1 clamp-on type transducer, pipe size from DN50-1000mm
TM
-1 clamp-on type transducer, pipe size from DN300-6000mm
TL
1.5 Typical application
The wall-mounting flow meter can be applied to a wide range of pipe flow measurements. Applicable liquids
include pure liquids as well as liquid with small quantity of tiny particles.
Examples are:
★ Water (hot water, chilled water, city water, sea water, waste water, etc.);
Sewage with small particle conte
Oil (crude oil, lubricating oil, diesel oil, fuel oil, etc.);
Chemicals (alcohol, acids, etc.);
Plant effluent;
Beverage, liquid food;
Ultra
-pure liquids;
Solvents and other liquids
1.6 Product Identification
Each set of the flow meter has a unique product identification number or ESN (electronic serial number)
written into the software that can only be modified with a special tool by the manufacturer. In
case of any hardware failure, please provide this number which is located on menu window M61 when
contacting the manufacturer.
1.7 Specifications
Items
Repeatability
Main unit
Measurement
Period
nt;
Accuracy
Better than ±1%
Better than 0.2%
Principle
Transit-time measuring principle
500ms
LCD with backlight, display accumulated flow/heat, instantaneous
Display
flow/heat, velocity, time etc.
Specifications
2