Connecting the AC Adapter
Connect the DC output of the AC adapter to the
connector indicated as "DC LINE" on the GL840.
Making Connections to the Analog Input Terminals
CH1...................................................CH20.
CAUTION: Connect wires to the specified channel, the channel number is shown on top of the terminal block.
Making Connections to the External Input/Output Terminals
* The B-513 input/output cable for GL (sold separately) is required for connecting input/output signals.
(For logic/pulse input, alarm output, trigger input, external sampling pulse input)
3
Connection Procedures
Resistance temperature detector input
+
-
b
Lead wire resistance should be
10 Ω or less per wire, three
wires need to be same length.
< Signal assignment >
Connecting the Grounding Cable
Use a flathead screwdriver to push the button above
the GND terminal while connecting the grounding
cable to the GL840. Connect the other end of the
cable to ground.
Voltage input
+
DC voltage input
-
A
+
B
-
b
b
Shunt resister
Ex: The current is converted to the voltage in the
shunt register.
For 4 to 20mA current input, installing 250 ohms
(0.1%) resister for converting 1 to 5V.
Note: 250 Ω shunt register is the option B-551.
Orange with red dotted line
Orange with black dotted line : 2
Grey with red dotted line
Grey with black dotted line
White with red dotted line
White with black dotted line
Yellow with red dotted line
Yellow with black dotted line : 4
Pink with red dotted line
Pink with black dotted line
Shielded
Thermocouple input
+
-
Compensation wire is
used if it is required.
Current input
+
-
: 1
Logic/pulse
: 3
input
: 4
: 1
: 2
Alarm output
: 3
: Trigger input/
external sampling input
GND