PARAMETER CHANGES

Tab
Flow Units
Flow Units?
Flow Units?
Flow Units
FSD(100% or URV) 200 l/m
Changing Selections
Note. Ensure that the correct access level is selected.
It is possible to enter data directly into the AquaMaster without navigating the menu system. This is
achieved by using the 'Command Line Interface' (CLI). To read the value of a variable, type
followed by the variable number (see overleaf), then press
To write a new value to a variable, type
by
. In both cases the AquaMaster will reply with the new value of the variable (or a reason for failure) e.g.:
>217
<0>217=42
>217=42
To change the flow FSD to 10 m
>248=setup
>112
>115
>115=10
>112=6
To cancel the password security, press
Space
: l/s
l/s
Space
l/m
: l/m
, the variable number,
>
<3>217=Write Access Denied
3
/min:
<0>248=2 logged in enter password
<0>112=1 l/s
<0>115=250
<0>115=10
3
<0>112=6 m
/m
Esc
(Repeat space function key until
required units are displayed)
Note. Use
input mode, or to exit a menu.
Where
217 is the number for the flow rate
<0> is the error code for 'no error'
42
is the current flow rate
<3> is the error code for 'Write Access Denied',
since writing to the flow rate is not allowed.
Comments
read present flow units
read present fsd
set new fsd value of 10
set units to number 6 (m
until the front screen is reached.
Esc
to escape or cancel in
.
and the new value required, followed
=
3
/min)
>
3