PARAMETER CHANGES

Tab
Esc
Note. Use
Changing Selections
Note. Ensure that the correct access level is selected.
It is possible to enter data directly into the AquaMaster without navigating the menu system. This is
achieved by using the 'Command Line Interface' (CLI). To read the value of a variable, type
followed by the variable number (see overleaf), then press
To write a new value to a variable, type
followed by
. In both cases the AquaMaster will reply with the new value of the variable (or a reason
for failure) e.g.:
Where:
>217
<0>217=42
>217=42
To cancel the password security, press
>248=setup
>112
>115
>115=10
>112=6
4
Flow Units
: l/s
Flow Units?
l/s
Flow Units?
l/m
Flow Units
: l/m
FSD(100% or URV) 200 l/m
to escape or cancel in input mode, or to exit a menu.
<3>217=Write Access Denied
Esc
<0>248=2 logged in enter password
<0>112=1 l/s
<0>115=250
<0>115=10
<0>112=6 m
Space
(Repeat space function key until
Space
required units are displayed)
, the variable number,
>
217 is the number for the flow rate
<0> is the error code for 'no error'
42
is the current flow rate
<3> is the error code for 'Write Access
Denied', since writing to the flow rate is not
allowed.
To change the flow FSD to 10 m
until the front screen is reached.
Comments
read present flow units
read present fsd
set new fsd value of 10
3
/m
set units to number 6 (m
.
and the new value required,
=
3
/min)
>
3
/min: