Set Mode - Extech Instruments SD750 User Manual

3-channel pressure datalogger
Table of Contents

Advertisement

Set Mode 
Before taking measurements and datalogging, the SD750 must be configured to match the 
application and transducers in question. The SET mode is used to accomplish the following: 
 
o Format a SD memory card  
o Set the clock date and time 
o Set the sampling time (datalogging capture rate) 
o Set the beeper sound ON/OFF 
o Set the meter to use a decimal (US) or a comma (EU) for readings stored on the SD card 
o Set the RS232 data output stream ON/OFF 
o Set CH1, CH2, and CH3 pressure sensor parameters to match the transducer(s) being used 
o Set the Pressure unit of measure (all channels use the same unit of measure) 
 
Note: The datalogger function must be switched OFF before attempting to enter SET mode. 
 
SET MODE MENU 
 
To enter the SET mode, press and hold the SET button for > 2 seconds, and then release. 'SET' 
will flash quickly on the display and the symbol for the first Menu item Sd F (format SD card) 
will appear. Follow the steps below to navigate the menu tree and to make and save changes 
to parameter settings. 
 
Note: In the SET mode, if no button is pressed within 5 seconds, the logger will exit the SET 
mode and return to the standard mode of operation. 
1.
Sd F: Format the SD card 
Use the ▲ button to select yES or no. To format (and completely erase) an SD card select 
'yES' and then press Enter. The formatting process will begin. To skip this step without 
formatting the card, select 'no' and then press SET to advance to the next parameter. 
2.
dAtE: Set the date and time 
Use the ▲ or ▼ buttons to adjust the Year/Month/Day/Hour/Minutes/Seconds. Use the 
Enter button to store and to step through the available fields. When done, Press SET to store 
the settings and to advance to the next parameter. 
3.
SP‐t: Set the sample time (datalogging capture rate) 
Use the ▲ button to select the desired sample rate. The options are: 5, 10, 30, 60, 120, 300, 
600 seconds and AUTO. In AUTO, a reading will be stored only when there is a value change of 
±10 digits. Press ENTER to store the setting and to advance. 
 