WEIGHBLOK WEIGHBLOK Digital Scale Setup & Operation Manual Revision 0502 4-2-03 ã 2001-2002 Transcell Technology, Inc. Contents subject to change without notice. Transcell Technology, Inc. 975 Deerfield Parkway Buffalo Grove, IL 60089 Tel (847) 419-9180 Fax (847) 419-1515 E-mail: transcell@transcell.net Web: www.transcell.net...
Page 2
The Weighingblock is a robust Scale. It is designed for portable use. When transported by car between different working places we recommend using of a protected space or a case. This will extend the interval for service. Manual- Weighblok Rev 1 4-2-03 975 Deerfield Parkway 800-503-9180 www.transcell.net...
Introduction Transcell new Weighblok series are a rugged digital scale casted in strong aluminum. It comes standard with a large (.75") LCDscreenfor easyreadout. All setup parameters may be entered via the membrane panel keys. The Scale has an ”intelligent” auto power off function and can be operated with 6 alkaline 1,5 V batteries.
Configuration The Weighblok has two menus. The Setup menu, containing most of the indicator's functional Setup parameters, consists of 15 separate menu selections, each with its own sub-menu of choices. The User menu, except A5 and A10 containing most of the indicator's serial communication parameters, which not are shown in this manual.
Allows the lb/kg key to be disabled and allows use of extended graduation. Disable the lb/kg "0" = Disable the lb/kg key "1" = Enable the lb/kg key ”2” Activate the extended graduation for VB2-100-EC Manual- Weighblok Rev 1 4-2-03 975 Deerfield Parkway 800-503-9180 www.transcell.net Buffalo Grove, IL 60089 847-419-1515 transcell@transcell.net...
Page 6
Selects the auto power off time period in minutes: 1, 2, 3, 5, 10, 15, 20, 30, off “Off” = Disabled (Always ON) Auto Power Off Period Manual- Weighblok Rev 1 4-2-03 975 Deerfield Parkway 800-503-9180 www.transcell.net Buffalo Grove, IL 60089 847-419-1515 transcell@transcell.net...
"Err2" – The internal resolution of the scale is not high enough to accept the calibration value. Select a larger parameter for the Span Gain (F2). Put the Calibration Switch to the normal mode. Check the Scale. Manual- Weighblok Rev 1 4-2-03 975 Deerfield Parkway 800-503-9180 www.transcell.net...
No reading from the ADC. Make sure there is a load cell(s) connected to the indicator at start-up. Err 9 Normal Operating Span calibration value has been lost. Re-calibrate the Scale. Mode Manual- Weighblok Rev 1 4-2-03 975 Deerfield Parkway 800-503-9180 www.transcell.net Buffalo Grove, IL 60089 847-419-1515 transcell@transcell.net...
Adjust and Calibrate the Scale. Bottom plate removal Remove the two screws from the bottom plate of the Weighblok.. Be careful with the keyboard connection. Check the inside of the Scale visually. Look for bad cables and connections. If the Scale shows F1 Check the position of the Calibration Switch.
TARE batteries. Indicates stable weighing. Weighblok Enunciator Definitions. Shown above is the LCD display. Before reading the value the stable indicator must be ”on”. If not, wait a few seconds until it appears. Basic scale use 1. Press ON to start the scale.
Communication Port The Weighblok has a Communication Port standard. It is possible to connect different accessories such as a Printer, PC/Notebook or external relay. Printer for the weight value. Many printers in the market can be connected. For more information, contact Transcell.