Recommended separation distances between portable and mobile RF communications equipment and the LMD-2735MD/
The LMD-2735MD/2435MD is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled.
The customer or the user of the LMD-2735MD/2435MD can help prevent electromagnetic interference by maintaining a minimum
distance between portable and mobile RF communications equipment (transmitters) and the LMD-2735MD/2435MD as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
W
0.01
0.1
1
10
100
For transmitters rated a maximum output power not listed above, the recommended separation distance in meters (m) can be
estimated using the equation applicable to the frequency of the transmitter, where is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Caution
When you dispose of the unit or accessories, you must
obey the laws in the relative area or country and the
regulations in the relative hospital regarding
environmental pollution.
Warning on power connection
Use a proper power cord for your local power supply.
1. Use the approved Power Cord (3-core mains lead) /
Appliance Connector / Plug with earthing-contacts
that conforms to the safety regulations of each
country if applicable.
2. Use the Power Cord (3-core mains lead) / Appliance
Connector / Plug conforming to the proper ratings
(Voltage, Ampere).
If you have questions on the use of the above Power
Cord / Appliance Connector / Plug, please consult a
qualified service personnel.
For the customers in U.S.A. and Canada
Warning on power connection for medical use
Please use the following power supply cord. With
connectors (plug or female) and cord types other than
those indicated in this table, use the power supply cord
that is approved for use in your area.
6
2435MD
Separation distance according to frequency of transmitter
150 kHz to 80 MHz
80 MHz to 800 MHz
= 1.2
0.12
0.38
1.2
3.8
12
Plug Type
Cord Type
Minimum Rating for Plug and
Appliance Couplers
Safety Approval
*Note: Grounding reliability can only be achieved when the equipment is
connected to an equivalent receptacle marked 'Hospital Only' or 'Hospital
Grade' .
Caution
When installing, the installation space must be secured
in consideration of the ventilation and service operation.
Leave a space 10 cm (4 inches) or more behind, 10 cm (4
inches) or more from the left and right sides of, 8 cm
1
(3
/
inches) or more from the bottom side of, and
4
7
30 cm (11
/
inches) or more above the unit.
8
For the customers in the U.S.A.
Caution
Federal law (United States of America)
restricts this device to sale by or on the order of
a licensed healthcare practitioner.
Warning
The apparatus shall not be exposed to dripping or
splashing. No objects filled with liquids, such as vases,
shall be placed on the apparatus.
Warning
To prevent injury, this apparatus must be securely
attached to the mounting arm/wall in accordance with
the instructions for use.
m
800 MHz to 2.5 GHz
= 1.2
= 2.3
0.12
0.23
0.38
0.73
1.2
2.3
3.8
7.3
12
23
United States and Canada
HOSPITAL GRADE*
Min. Type SJT
Min. 18 AWG
10A/125V