CLEAR CALIBRATION

Press Clear functional key when displayed to clear old calibrations.
All old calibrations will then be cleared and the instrument continues calibration. The points confirmed
in current calibration are kept.
Note: If Clear calibration is invoked during the first calibration point, the instrument returns
to measurement mode.

REPLACE CALIBRATION STANDARD

Each time a standard is confirmed the new calibration parameters replace the old calibration parameters
of the corresponding standard.
If the current standard has no correspondence in the existing stored calibration structure and this is not
full, the current standard is added to the existing stored calibration (up to 5 standards).
If the existing stored calibration is full (five calibration points) after confirming the calibration point,
the instrument will ask which buffer will be replaced by current standard.
Press
/ keys to select another standard to be replaced.
Press CFM to confirm the standard that will be replaced.
Press CAL or ESC to leave replace mode. In this case, the standard will not be stored.
Note: The replaced standard is not removed from calibration list and it can be selected for
the next calibration points.
31