Temperature Calibration - Hanna Instruments HI98192 Instruction Manual

Usp compliant ec, tds, nacl, resistivity, temperature meter
Table of Contents

Advertisement

• Press CFM to confirm.
• The second expected calibrated point is displayed.
• Immerse the probe into the second vessel as close as possible to the reference thermometer. Allow
a few seconds for the probe to stabilize.
• Use the / keys to set the calibration point value to that measured by the reference
thermometer.
• When the reading is stable and within range of the selected calibration point, CFM functional key
is displayed.
• Press CFM to confirm. The instrument returns to measurement mode.
54