GENERAL PARAMETER SCREENS

Select Profile

Highlight Select Profile.
Press Select.
The list with stored profiles is displayed.
Press Add to add a new profile to the list (max 10).
Use the / keys to highlight the desired profile.
Press Select to choose the profile and exit to SETUP.
Press View to view profile information.
The profile information includes the date and the time when the profile was added, information about
calibration on EC and NaCl range, the setup cell constant, information about temperature setting,
reference temperature, temperature compensation mode, temperature coefficient, temperature source
and information about fixed ranges.
If calibration exists, the GLP functional key is displayed. Press GLP EC to view the corresponding
GLP EC information.
Press GLP NaCl to view the corresponding GLP NaCl information.
Note: If the EC or NaCl calibration was not performed while the current profile is set, or
calibration was cleared, the corresponding GLP key is not displayed.