Olimpus CKX41 Instruction Manual page 32

Culture microscopes
Table of Contents

Advertisement

²
@
@
|
³
²
@
²
28
Fig. 31
@
Fig. 32
Mounting the CKX-TBI Tilting Observation Tube
}The tube mount | needs to be removed only when using this observation tube.
(After removal, retain the tube mount and clamping screw carefully for use in mount-
ing another observation tube later.)
1. Loosen and remove the tube-clamping knob @.
2. Peel off the sticker ² hiding the tube mount clamping screw.
3. Loosen the clamping screw ³ and remove the tube mount | from the microscope
frame.
4. Further loosen and remove the clamping screw ³ and attach the tube-clamping
knob @ in the same position (be sure to store the clamping screw ³).
5. Fit the round dovetail at the bottom of the observation tube into the dovetail on the
microscope frame, correct the tube position so that the binocular section faces the
operator side, and tighten the tube clamping knob @.
Fig. 33
6
Mounting the Stage Center Plate (CKX41 Only)
Fit the standard stage center plate 1 into the opening on the stage.
}Turn the center plate so that the notch 2 faces to the front for easy
confirmation of the objective tip.
# When using the glass stage center plate, set it in the direction
so that its product code engraving (CK40-CPG) can be read from
the front.
7
Mounting the Observation Tube (CKX41 Only)
1. Loosen the observation tube clamping knob 1 to a degree at which the
knob does not come out. (Fig. 32)
2. Fit the circular dovetail at the bottom of the observation tube into the
observation mount on the microscope frame. Adjust the observation tube
until the binocular eyepieces face directly to the front, and then tighten
the clamping knob. (Fig. 32)
8
Mounting the Eyepieces
Insert the eyepiece 1 into each eyepiece sleeve 2 on the observation
tube.
(Fig. 31)
(Fig. 32)
(Fig. 33)

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for Olimpus CKX41

This manual is also suitable for:

Ckx31

Table of Contents