³
|
²
18
@
²
5
Fig. 21
@
Fig. 22
2
Mounting the Optical Element
}When the IX2-SL phase contrast slider is used, it is required to attach an
optical element matching the phase objective in use.
1. Hold the IX2-SLPH2 optical element 1 so that engraving faces upward
and drop in the optical element in the empty position 2.
2. When mounting the IX2-SLPHC or IX2-SLPH1 optical element 3, insert
the provided centering knobs into the screw holes 5, loosen them fully
before dropping the optical element in the empty position 4 so that the
engraving faces upward.
3. After dropping in the optical element, lightly tighten the centering knobs.
3
Mounting the Phase Slider
1. Hold the phase slider 1 face up (engraving side up) with the finger hold
on the right, and insert it into the illumination column slot.
2. When performing phase contrast observation, always set the aperture iris
diaphragm lever 2 to "(" (wide open).
(Fig. 21)
(Fig. 22)