9-2 Detailed Assembly Procedures
²
³
5
³
|
26
@
²
³
Fig. 25
|
@
Fig. 26
@ ²
Fig. 27
1
Installing/Replacing the Halogen Bulb
}Use only the designated high-intensity halogen bulb 6V30WHAL (PHILIPS
5761).
To prevent reduced bulb life or cracking, do not touch the bulb with
bare hands. If fingerprints are accidentally left on the bulb, wipe the
bulb with a soft cloth.
· Holding the bulb 1 with gloves or a piece of gauze, insert the bulb pins
2 fully into the pin holes 3 on the lamp socket.
# Insert the bulb gently. Squeezing too hard will damage the bulb.
Caution for Bulb Replacement During Use or Right After Use
The bulb, lamp socket and areas near these will be extremely hot
during and right after use.
Set the main switch to "
the wall outlet, then allow the old bulb and lamp socket to cool
before replacing the bulb with a new of the designated type.
2
Installing the Transmitted Light Illuminator
1. While aligning the index groove 2 on the transmitted light illuminator 1
with the notch 3 on the illumination column, insert the transmitted light
illuminator 1 gently into the illumination column.
2. Rotate the transmitted light illuminator 1 90° clockwise so that letters
"AS" 4 on the filter holder faces directly the front. Then tighten the clamping
screw 5 using the Allen wrench provided with the microscope frame.
3
Installing the Lamp Socket
· Connect the plug 1 to the socket pin 2. The, while aligning the guide
pins 3 with the condenser's guide holes 4, push the lamp socket gently
into the transmitted light illuminator.
" (OFF), disconnect the power cord from
(Fig. 25)
(Fig. 26)
(Fig. 27)