USING THE CONTROLS

4-1 Microscope Frame

@
Fig. 3
@
Fig. 4
@
Fig. 5
1

Turning On the Light Source

Set the main switch 1 on the side panel of the microscope frame to " I "
(ON).
2

Adjusting the Brightness

Turn the light intensity control 1 clockwise to raise the voltage and increase
the brightness. Turn it counterclockwise to lower the voltage and decrease
the brightness.
}The service life of the bulb can be extended by using the bulb at a lower
voltage.
Adjusting the Tension of the Coarse
3
Adjustment Knob
# Be sure to use the tension adjustment ring 1 to adjust the rotation
tension of the coarse adjustment knob.
Adjustment Procedure
Turn the tension adjustment ring 1 with your fingers or using a flat-blade
screwdriver. When the ring is turned in the direction of the arrow, tension
of the coarse adjustment knob increases. Turning the ring in the opposite
direction decreases the tension.
If the revolving nosepiece descends on its own or if the specimen gets
out of focus quickly even when it is brought into focus using the fine
adjustment knob, it means that the tension of the coarse adjustment
knob is too low. Turn the ring in the direction of the arrow to increase the
tension.
CKX41/CKX31
(Fig. 3)
(Fig. 4)
(Fig. 5)
9