Trouble
f ) Picture is too dark or bright.
g) The colors in the picture are
strange.
h) Error message is displayed
during data downloading to
PC.
}If no operation can be activated by pressing any of the buttons, unplug the AC adapter's power cord from the power outlet,
plug it again and set the main switch to " I " (ON) again.
40
Cause
AE LOCK, EXP.ADJ. and/or SPOT are not
set properly.
The illumination brightness is not set
properly.
A fluorescent lamp is in use.
A wrong color temperature is set in the
white balance control operation.
The PC connection cable is not con-
nected properly.
The main switch of the control box is OFF.
Remedy
Set them properly.
Adjust the brightness.
Use an illumination other than a fluores-
cent lamp.
Set the color temperature properly.
Connect it properly.
Set the main switch to " I " (ON).
Page
17/18
21
33
13