8.5 Interface RS232C

Data output via interface RS 232 C
General Information
The previous condition for the data transfer between balance and a peripheral device
(e.g. printer, PC ...) is that the appliances are set to the same interface parameters
(e.g. baud rate, transfer mode ...).

8.5.1 Data transfer mode

Setting of data transfer mode
1. Keep the
key pressed until „Unit" is displayed.
2. Press
key, „Pr" appears.
3. To change the setting, press the
4. The mode can be converted using the
(Pr PC; AU PC; AU Pr ; re Cr ; BA Pr ;details chapter 9.4).
5. The changed setting is taken over by pressing
the
key.
6. The balance returns to weighing mode.

8.5.2 Baud rate

The baud rate necessary for the measuring values transfer can be set.
In the following example the baud rate is set to 9600 baud.
Setting of baud rate
1. Keep the
key pressed until „Unit" is displayed.
2. Press
key
3. Press
key, „baud" appears.
4. Confirm by pressing the
5. The baud rate can be changed using the
(1200, 2400, 4800, 9600, 19200).
6. The changed setting is taken over by pressing the
7. The balance returns to weighing mode.
64
key.
key
key
Balance display
Unit
Pr
Pr PC
AU Pr
0.0 g
Balance display
Unit
Pr
Baud
4800
key
9600
key.
0.0 g
DE-BA-defsi-0640