Table 5-1. Calibration Component Identification - National Instruments SC-2040 User Manual

Eight-channel simultaneous sample-and-hold accessory
Table of Contents

Advertisement

Chapter 5
Your accuracy needs determine how carefully the offsets need to be calibrated. A typical
requirement might be for total offset referred to output to be less than half of an LSB of the DAQ
board being used. For example, a 12-bit, 20 V system has a resolution of 20 V/2
Calibration to under 2 mV would thus be sufficient for most applications. The SC-2040 is
factory calibrated to have total offset referred to output of less than 2 mV at low gains.
Table 5-1 shows which pots to adjust for each type of offset for each channel and which DIP
switch selects the gain for each channel. Refer to the parts locator diagram in Chapter 2,
Configuration and Installation, to determine the location of each component.
Channel
Output Offset Adjust
0
1
2
3
4
5
6
7
A complicating factor in the calibration of the board is that the output offset is not the same in
track mode as in hold mode. This difference is because of a phenomenon known as hold step, in
which a small amount of charge is transferred to the track-and-hold amplifier hold capacitor
during the transition from track mode to hold mode. This charge transfer slightly changes the
voltage at the output of the track-and-hold amplifier.
The magnitude of the hold step on the SC-2040 is approximately 5 mV, enough to make it
necessary to calibrate output offset in hold mode rather than track mode. You may calibrate
input offset without switching the board to hold mode.
To calibrate the SC-2040 board, the following steps are required:
1. If you are using a rack-mount kit, remove the cover to expose the DIO switches and jumpers.
Connect the board so that the pots and DIP switches are readily accessible.
2. Short each input of the board to ground.
At low gains, more alternatives exist to serve as a short circuit, but at a gain of 500, 1 mV
referred to the output is only 2 V referred to the input. Thermoelectric effects make it
difficult to keep a short circuit from generating such voltages. The best solution is to use
short lengths of heavy copper wire, and to keep the short away from heat sources.
3. Program the DAQ board to take data in the differential input mode.
The SC-2040 should switch between track mode and hold mode, with the DAQ board
sampling occurring while the board is in hold mode.
© National Instruments Corporation

Table 5-1. Calibration Component Identification

R1
R7
R13
R18
R25
R32
R37
R45
Input Offset Adjust
R4
R11
R16
R22
R29
R35
R41
R48
5-3
Calibration Procedures
12
= 4.88 mV.
Gain DIP Switch
U2
U6
U9
U14
U17
U21
U24
U28
SC-2040 User Manual