Specifications (cont'd)
Pollution Degree: 2
Battery: 9 V battery (NEDA 1604, JIS 006P or IEC 6F22)
Electric shock hazard:
Products covered by this manual were originally equipped with a different
fuse in location FS1. These units can be identified by the TUV/GS logo on
the front of the unit. For units bearing this logo, the following specifications
apply. All other specifications are unchanged.
Overload Protection:
µA and mA: (FS1) 1 A/600 V fast acting fuse, IR 100 kA, 13/32" x 1-1/2"
Measurement Category:
µA and mA Terminal: Category III and IV, 600 VAC and 300 VDC
All Category IV ratings were in accordance with IEC 61010-1 2nd edition draft (1999).
Do not exceed these ratings. The 1000 V fuses specified below for FS1 may be used in
these units, but the measurement category for these meters is unchanged.
Failure to observe this warning could result in severe injury or death.
Overload Protections:
µA and mA: (FS1) 1 A/1000 V fast acting fuse, IR 10 kA, 13/32" x 1-1/2"
(or 0.44 A/1000 V fast acting fuse, IR 10 kA, 13/32" x 1-3/8")**
Amps: (FS2) 15 A/1000 V, IR 10 kA (or 11 A/1000 V, IR 20 kA) fast acting fuse,
13/32" x 1-1/2"
Volts, Ω, and Others: 1450 V peak, 1050 VAC RMS
Measurement Category:
Category III, 1000 V per UL 61010B-1
Category IV, 600 V per IEC 61010-1, 2nd edition
** The 0.44 A fuses designed for DMM protection (Bussman DMM-B-44/100 and Littelfuse
FLU.44) will carry 500 mA for thousands of hours.
DM-860
23