Greenlee DM-860 Instruction Manual page 18

Hide thumbs Also See for DM-860:
Table of Contents

Advertisement

Accuracy
See the Specifications section for operating conditions and temperature coefficient.
Accuracy is specified as follows: ± (a percentage of the reading + a fixed amount)
at 23 °C ± 5 °C (73.4 °F ± 9 °F), 0% to 75% relative humidity. Specifications are for 50,000
counts mode.
True RMS Readings: True RMS voltage and current accuracies are specified from 5% to 100%
of the range unless otherwise specified. Frequency must be within the specified bandwidth for
non-sinusoidal waveforms. Crest factors are as follows:
• Crest factor < 5:1 at full scale
• Crest factor < 10:1 at half scale
AC and DC + AC Current
Range
500.00 µA
5000.0 µA
50.000 mA
500.00 mA
5.0000 A
10.000 A*
*10 A continuous; 20 A maximum (Duty Cycle: 30 seconds on, 5 minutes off)
Range
500.00 µA
5000.0 µA
50.000 mA
500.00 mA
5.0000 A
10.000 A*
18
Accuracy at
50 Hz to 60 Hz
± (0.5% + 0.5 µA)
± (0.5% + 5.0 µA)
± (0.5% + 0.05 mA)
± (0.5% + 0.5 mA)
± (0.5% + 0.005 A)
± (0.5% + 0.05 A)
Burden Voltage
(All Frequency Ranges)
0.15 mV/µA
0.15 mV/µA
3.3 mV/mA
3.3 mV/mA
0.03 V/A
0.03 V/A
Accuracy at
40 Hz to 1 kHz
± (0.7% + 0.5 µA)
± (0.7% + 5.0 µA)
± (0.7% + 0.05 mA)
± (0.7% + 0.5 mA)
± (0.7% + 0.005 A)
± (0.7% + 0.05 A)
Accuracy at
1 kHz to 10 kHz
± (2.0% + 0.5 µA)
± (2.0% + 5.0 µA)
± (2.0% + 0.05 mA)
± (2.0% + 0.5 mA)
Unspecified
Unspecified

Advertisement

Table of Contents
loading

Table of Contents