Error 1
Sample Application:
Cause: Meter detected problem with color of
test strip.
Solution:
1.Repeat test with new test strip and
• Apply blood to pink test square only.
• Make sure confirmation dot turns completely
blue.
• Insert test strip within 2 min of applying blood.
2.Check that sample hematocrit is within range:
adult 25–60%, neonatal 25–65%.
3.If you continue to get any error message,
obtain a laboratory result, and
• Compare color of test strip confirmation dot
to Color Chart on test strip bottle to see if
glucose level is very high or very low.
• Take appropriate steps to monitor patient
and provide medical care while awaiting
laboratory result.
4.Clean meter and repeat test with new test strip.
5.Contact LifeScan for help at 1 800 524-7226.