Control solution testing
OneTouch
®
Ultra
check that the meter and the test strips are working properly.
Do a control solution test:
• to practice the test process instead of using blood,
• once a week,
• whenever you open a new vial of test strips,
• if you suspect the meter or test strips are not working properly,
• if you have had repeated unexpected blood glucose results (as described on page 25), or
• if you drop or damage the meter.
NOTE:
• Use only OneTouch
• Control solution tests must be done at room temperature (20–25°C). Make sure your
meter, test strips, and control solution are at room temperature before testing
28
®
Control Solution contains a known amount of glucose and is used to
®
Ultra
®
Control Solution with your OneTouch
®
UltraEasy
™
Meter
.
.