Invacare IRC5PAW Service Manual page 83

Perfecto2 oxygen concentrator (homefill ii compatible)
Hide thumbs Also See for IRC5PAW:
Table of Contents

Advertisement

@ 1698675  76987 94 75/27 5 7698742679754752 5 75794 4 52947
6 5 673 5 7547BAH776 7547D@H7$224761724 27 7?>>$554 67
24 259 7623782667266245 754 7 7427826679424 797 4
A 65797 7-*D,2479/3 47 4 9 72437 8 47 79/3 47 4 9 754 7 7
 9 75487 3659 
4 75 74979779/3 47 4 9 724378 47652437 79/3 47 4 9 72437
. $224761724 27 7/3 471 4 9 765 7 96 75255 754 7 79754 676237
24 259 782667266245 7" 79 9 782667 7 9847827 7565 
LABEL
SYMBOL
SYSTEM OKAY
O
2
O
over 85% (± 2%)
2
O
Between 73% (± 3%)
2
to 85% (±2%)
SYSTEM FAILURE
O
Below 73% (±3%)
2
Part No 1154245
87
FIGURE 24.3 Oxygen Sensor
SENSO
ALARM THRESHOLD
2
STATUS
GREEN Indicator Light
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor Failure
Call a qualified technician.
RED Indicator light
Continuous Audible Alarm Sieve-GARD™Compressor
Shutdown
Call a qualified technician
83
SECTION 24—ALARM TEST
1/8-inch I.D. Oxygen Sensor Tubing
(Clamp Off Flow)
INDICATOR LIGHTS