Invacare IRC5PAW Service Manual page 17

Perfecto2 oxygen concentrator (homefill ii compatible)
Hide thumbs Also See for IRC5PAW:
Table of Contents

Advertisement

SYMPTOM
Power Loss:
(Continued)
Internal Status Indicators:
RED: Off
GREEN: Off
Unit plugged in, power
switch, alarm off. Battery
Free circuit drained.
Internal Power
Loss Senso
:
2
Internal Status Indicators:
RED: Off
GREEN: Off
Alarm may or may not be on.
Control Panel Indicators:
RED: Off
YELLOW: Off
GREEN: Off
Fan operates, Compressor
not operating.
High Pressure:
Internal Status Indicators:
RED: One Flash
GREEN: Two Flashes
Unit plugged in, power switch
on, continuous alarm.
Compressor shut down.
Part No 1154245
PROBABLE CAUSE
P.C. board.
a. P.C. board damaged.
b. Loose or damaged connector.
Transformer assembly.
a. Faulty.
b. Connector loose or
disconnected.
c. Faulty wiring.
P.C. board faulty.
P.C. board:
a. Malfunction
b. Disconnected wire.
c. Shifting valve at pressures greater
than 25 p.s.i. for Perfecto
4-way Valve:
a. Not shifting.
b. Bad coil.
c. Coil resistance.
17
SECTION 4—TROUBLESHOOTING
SOLUTION
a. Replace P.C. Board. Refer to
Replacing P.C. Board on page 58.
b. Repair or replace connector.
a. Replace. Refer to Replacing the
Transformer on page 61.
b. Reattach connector.
c. Replace transformer assembly. Refer
to Replacing the Transformer on
page 61.
Replace P.C. board. Unit requires retiming
after P.C. board replacement. Refer to
Replacing P.C. Board on page 58.
a. Set flow to max L/min. for
concentrator. Check voltage across
Pilot Valve 1 on volt DC scale. If
meter reads 0 volts when unit is
turned on, replace P.C. board
b. Check spade connectors on pilot
valves 1 or 2 and connectors on P.C.
board.
c. Replace P.C. board. Unit requires
retiming after P.C. board
.
replacement. Refer to Replacing P.C.
2
Board on page 58.
a. Check voltage. If voltage rises to
approximately 24 volts D.C. when
unit shuts down, P.C. board is
functioning properly. Replace the
valve if shorted or open. Refer to
Replacing 4-Way Valve and/or
Manifold Assembly on page 65.
b. Replace 4-way valve. Refer to
Replacing 4-Way Valve and/or
Manifold Assembly on page 65.
c. Coil should read 175 ohms ± 20.
Perfecto2™