Device messages

Device message/possible cause
High leak detected, connect your tubing
Air tubing may not be connected properly.
Mask may be fitted incorrectly.
Tubing blocked, check your tubing
Air tubing may be blocked.
SD card error, remove your card and press Start to begin therapy
SD card may not be inserted correctly.
Read only card, please remove, unlock and re-insert SD card
SD card switch may be in the lock (read-only) position.
System fault, refer to user guide, Error 004
Device may have been left in a hot environment.
Air filter may be blocked.
Air tubing may be blocked.
There may be water in the air tubing.
All other error messages, for example, System fault, refer to user guide, Error 0XX
An unrecoverable error has occurred on the device.
10
Solution
Make sure the air tubing is firmly connected at both ends.
Make sure your mask is fitted correctly. See your mask user
guide for fitting instructions or use the Mask Fit function to
check your mask fit and seal.
Check the air tubing and remove any blockages. Press the
dial to clear the message and then press Start/Stop to
restart the device.
Remove and reinsert the SD card.
Move the switch on the SD Card from the lock position
the unlock position
and then re-insert it.
Allow to cool before re-use. Disconnect the power supply
and then reconnect it to restart the device.
Check the air filter and replace it if there are any blockages.
Disconnect the power supply and then reconnect it to restart
the device.
Check the air tubing and remove any blockages. Press the
dial to clear the message and then press Start/Stop to
restart the device.
Empty the water from the air tubing. Disconnect the power
supply and then reconnect it to restart the device.
Contact your care provider. Do not open the device.
to