AirSense 10 and AirCurve 10 reprocessing guide
This guide is intended for multipatient use of the AirSense
When the device is used for multiple patients, for example, in a sleep lab, clinic, hospital or at a health care provider, the
cleanable water tub, air outlet and air tubing should be reprocessed between each patient use.
If the cleanable water tub or the air tubing are being used for a single user in the home, refer to the cleaning instructions in the
User Guide.
Described here are ResMed's recommended and validated procedures for cleaning and disinfecting the cleanable water tub, air
outlet and air tubing. However, the steps for disinfection vary regionally and each healthcare facility should consult its own
procedures before carrying out those within this guide.
Note: The standard water tub cannot be disinfected. If contaminated, it must be discarded and replaced with a new water tub.
WARNING
 ResMed cannot give any assurance that deviations from the procedures listed in this guide, and their effect on the
performance of the product, will be acceptable.
 When using detergents, disinfectants or sterilization agents, always follow the manufacturer's instructions.
 Beware of electrocution. Do not immerse the device, power supply or power cord in water. If liquids are spilled into
or onto the device, unplug the device and let the parts dry. Always unplug the device before cleaning and make sure
that all parts are dry before plugging it back in.
Surface disinfection
1. Wipe the exterior of the device including display, externally accessible ports, side cover, power supply unit and accessories
with a disposable cloth and mild detergent or alcohol disinfectant (see list below).
2. Remove any excess disinfectant with a disposable dry cloth.
Agents recommended for surface disinfection and cleaning:
Warm water and mild detergent eg, Teepol
Window cleaner or other premixed surface detergent
Methyl alcohol solution
70% Ethyl alcohol solution
70-90% Isopropanol solution
10% Bleach solution
Isopropyl wipes
CaviCide
Mikrozid
®
Actichlor
Plus
Terralin
®
.
Note: Agents may not be available in all regions.
10 and AirCurve
multipurpose detergent
10 devices.
 
English
English
1

Advertisement

loading

Summary of Contents for ResMed AirSense 10

  • Page 1 Note: The standard water tub cannot be disinfected. If contaminated, it must be discarded and replaced with a new water tub. WARNING  ResMed cannot give any assurance that deviations from the procedures listed in this guide, and their effect on the performance of the product, will be acceptable.
  • Page 2 2. Run the detergent solution through the air tubing repeatedly until no contamination is visible. 3. Thoroughly rinse each component according to the detergent manufacturer's instructions. ResMed has tested the following detergents according to the manufacturer’s instructions: Detergent Water temperature...
  • Page 3   High level chemical Part Validated number of cycles CIDEX ® OPA Ortho-phthalaldehyde 0.55% for 12 minutes Gigasept FF ® (new) 5% for 15 minutes SlimLine ClimateLineAir Standard 1. Soak the air tubing in a commercially available solution of a chemical sterilant. Take care that no air bubbles are trapped inside the air tubing.
  • Page 4 1. Clean all components with a soft bristled brush for one minute while soaking in detergent solution (see table below). Pay particular attention to all crevices and cavities. 2. Thoroughly rinse each component according to the detergent manufacturer's instructions. ResMed has tested the following detergents according to the manufacturer’s instructions: Detergent Water temperature...
  • Page 5   Disinfecting In the procedures below, only one disinfection process needs to be performed. High level thermal Part Validated number of cycles Hot water: 194°F (90°C) for 1 minute. Cleanable water tub Air outlet 1. Soak the disassembled components in a hot water bath at 194ºF (90ºC) for 1 minute. Take care that no air bubbles are trapped against the components.
  • Page 6   Reassembling The following instructions provide guidance on how to correctly reassemble the air outlet and the water tub. To reassemble the air outlet 1. Hold the air outlet with the seal pointing to the left and the clip pointing forward. 2.
  • Page 7: Packaging And Storage

    See www.resmed.com for other ResMed locations worldwide. For patent and other intellectual property information, see www.resmed.com/ip. Air10, AirSense, AirCurve, ClimateLine, HumidAir and SlimLine are trademarks of ResMed Ltd. ClimateLine and SlimLine are registered in U.S. Patent and Trademark Office. Actichlor is a trademark of Ecolab US Inc. Alconox is a trademark of Alconox Inc.

This manual is also suitable for:

Aircurve 10