Electromagnetic Compatibility
Product complies with all applicable electromagnetic compatibility requirements
(EMC) according to IEC60601-1-2, for residential, commercial and light industry
environments. For further details, see "Guidance and Manufacturer's
Declaration - Electromagnetic Emissions and Immunity" on page 40.
Air Filter: Two-layered, powder-bonded, polyester non-woven fiber
Air Tubing: Flexible plastic, 6'6" or 9'10" length
IEC 60601-1 Classifications
Class II (double insulation)
Type CF
Continuous operation
This flow generator is not suitable for use in the presence of a flammable
anesthetic mixture with air, or with oxygen or nitrous oxide entrained in the flow
generator airpath.
Table 7: Displayed values
Value
Pressure sensor at air outlet
Pressure
Flow sensor in flow generator*
Leak
Tidal volume
Respiratory rate
Minute ventilation
*
Results may be inaccurate in the presence of leaks.
For tidal volume up to 800 mL.
For minute ventilation up to 10 L/min.
Note: The manufacturer reserves the right to change these specifications
without notice.
38
Range
Accuracy
-5 to 30 cm H
O
±0.5 cm H
2
0–120 L/min
Greater of ±6 L/min
or ±20%
100–3000 mL
±5%
6–60 BPM
±0.5 BPM
0.6–60 L/min
±7%
Display Resolution
O
0.1 cm H
2
1 L/min
1 mL
1 BPM
0.1 L/min
O
2