HP 3478A Technical Manual page 61

Dilgital multimeter
Hide thumbs Also See for 3478A:
Table of Contents

Advertisement

Hewlett-Packard Model 3478A
Digital Multimeter
Serial Number_____________
Ohms Test
Step#
Input to
3478A
1
Open
2
Short
3
Short
4
Short
5
Short
6
Short
7
Short
8
Short
9
30 ohm
(9)
10 ohm
10
300 ohm
(10)
100 ohm
11
3K ohm
(11)
1K ohm
12
30K ohm
(12)
10K ohm
13
300K ohm
(13)
100K ohm
14
3M ohm
(14)
1M ohm
15
30M ohm
(15)
10M ohm
PERFORMANCE TEST CARD
1 YEAR LIMITS
Set-Up and
Configuration
Press TEST/RESET
Ohms Function
00.0041
300 Range
000.005
3K Range
0.00002K
30K Range
00.0002K
300K Range
000.002K
3M Range
0.00002M
30M Range
00.0002M
30 Range
30.0122
10.0068
300 Range
300.041
100.017
3K Range
3.00035K
1.00013K
30K Range
30.0035K
10.0013K
300K Range
300.035K
100.013K
3M Range
3.00035M
1.00013M
30M Range
30.0200M
10.0068M
9/(10 blank)
Test Performed By______________________
Date_____________
Reference Temperature__________________
High
Reading
Limit
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
TM 11-6625-3071-14
Low
Test
Limit
Pass
00.0000
_____
000.000
_____
0.00000K
_____
00.0000K
_____
000.000K
_____
0.00000M
_____
00.0000M
_____
29.9878
_____
09.9932
_____
299.958
_____
099.983
_____
2.99965K
_____
0.99987K
_____
29.9965K
_____
09.9987K
_____
299.965K
_____
099.987K
_____
2.99965M
_____
0.99987M
_____
29.9800M
_____
09.9932M
_____
Test
Fail
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents