HP 3478A Technical Manual page 92

Dilgital multimeter
Hide thumbs Also See for 3478A:
Table of Contents

Advertisement

3478A
Figure 7-D-7. JM403 SA Connection
4. If pin 3 of U466 is high, replace U466. If low,
replace U467.
c. If the RESET line is high, while monitoring the
line, press the 3478A's blue Shift button and then the
SGL/TRIB (TEST/RESET) button. The 3478A should go
through its Self-Test routine, and when the routine is
finished, the RESET line toggles from high to low to high.
d. If the RESET line toggles and the 3478A is still
in- operative, replace U462.
e. If the RESET line does not toggle, do the
following:
1. Make sure pin 1 of U467 is toggling.
2. If pin 1 is not toggling, the Isolation Cir- cuitry
may be at fault.
troubleshooting.
3. If pin 1 is toggling, replace U486 and then U467.
7-D-44. ISOLATION CIRCUITRY TROUBLESHOOTING
7-D-45. The Isolation Circuitry transfers information from
the Main Controller (U501) to the A/D Controller (U462)
and also from the A/D Controller to the Main Controller.
A failure in the logic is normally indicated when, after
doing a Self-Test, "A:D LINK FAIL" is displayed.
troubleshoot the circuitry, first determine the inoperative
circuitry and then troubleshoot the cir- cuitry.
following:
a. Turn the 3478A off.
b. Move jumpers JM502 and JM503 to the "D"
position.
c. Connect the Floating Common ground to the
3478A's chassis. The LO INPUT terminal can be
Go to paragraph 7-D-44 for
used as a Floating Common ground (make sure the cor-
rect terminal is used, dependent on the position of the
Front/Rear Switch).
d. Connect and set the Signature Analyzer as
follows:
Start:
Stop:
Clock:
Hold:
Self-Test: Out
Gnd:
e. Turn the 3478A on and check the signature of
the Chassis Common + 5V power supply.
f.
SA probe is toggling, the Main Controller (U501) may be
defective. Go to Flowchart A for troubleshooting (see
paragraph 7-D-II 1).
g. If the SA probe is not toggling (the signature can
be the same or different), the isolation circuitry is defec-
tive. Do the following:
U508 are toggling.
transformer T501 and associated circuitry. If T501 is
good, replace U468.
h. If the signals at pin 38 of U501 and pin 14 of
U462 are good, the isolation logic used to transfer
information between U462 and U501 may be defective.
Do the following:
is toggling.
To
tive. Go to Flowchart D for troubleshooting.
Do the
U467 are toggling.
transformer T401 and associated circuitry. If T401 is
good, replace U550.
7-D-10
U462 pin 1 (
)
TP8 (
)
TP3 (
)
Out
Chassis Ground
If the signature is different than "HF52" and the
1.Using a logic probe, make sure pins 2 and 4 of
2.If the pins are not toggling, replace U508.
3.If they are toggling, check for a defective
1.Using a logic probe, make sure pin 38 of U462
2.If pin 38 is not toggling, U462 may be defec-
3.If pin 38 is toggling, make sure pin 1 and 4 of
4.If the pins are not toggling, replace U467.
5.If they are toggling, check for a defective
TM 11-6625-3071-14