Ohms Test.4-10; Ac Current Accuracy Test And Calibration; 4-Wire Ohms Connection; Ohms Test Limits - HP 3478A Technical Manual

Dilgital multimeter
Hide thumbs Also See for 3478A:
Table of Contents

Advertisement

3478A
Figure 4-5. AC Current Accuracy Test and Calibration
AC-DC Current Calibrator to the IA Range. Check and
record the 3478A's reading.
j. Remove the AC Calibrator from the AC-DC Current
Calibrator.
Remove the current calibrator from the
3478A. This completes the AC Current Test. If any test
fails, try calibrating the instrument (go to paragraph 4-
94). If the test still fails, go to this manual's Section VII
for troubleshooting.
4-69. Ohms Test
4-70. The Ohms Test should only be done in the ohms
function in which the 3478A was last calibrated. If, for
example, the 3478A was last calibrated in the 4-Wire
Ohms Function, perform the Ohms Test in the 4-Wire
Ohms Function. If the 3478A was last calibrated in the
2-Wire Ohms Function, perform the Ohms Test in the
3478A
2-Wire Ohms Function.
connections or test leads used to calibrate the 3478A in
the 2-Wire Ohms Function must also be used when
testing the multimeter in that function. This is because
the impedance of the test leads is canceled out when the
multimeter is calibrated (in the 2-Wire Ohms Function). A
different set of test leads can have different impedances
which show up as an ohms test error.
4-71. The Ohms Function of the 3478A can be tested
using either full scale or 1/3 scale inputs (e.g. 3K ohm or
1K ohm resistors to check the 3K ohm Range). Because
of this, both test limits are in Table 4-7 and on the Ohms
Performance Test Card. The test step numbers (on the
Test Card) shown in parenthesis are for the 1/3 scale
tests. When testing the 3478A, make sure the test leads
used to connect the standard resistor to the 3478A's
INPUT Terminals are as short as possible. This is to
prevent the leads from picking up any noise, which could
give an invalid reading.
4-72. If the 3478A is to be tested in the 2-Wire Ohms
Function, all test resistors are applied to the 3478A's
INPUT Terminals. If the 4-Wire Ohms Function is to be
tested, short the INPUT Terminals and connect the test
resistors to the terminals, as shown in Figure 4-6 and
Figure 4-7, respectively.
4-73. The Ohms Test limits are printed on the Ohms
Test Card and in Table 4-7. Each step on the test card
corresponds to a step in the procedure with each step
number shown in parenthesis in the procedure.
4-74.
required test equipment for the Ohms Test.
Table 4-7. Ohms Test Limits
4-10
In addition, the same test
Equipment Required.
TM 11-6625-3071-14
The following is the