HP 3478A Technical Manual page 59

Dilgital multimeter
Hide thumbs Also See for 3478A:
Table of Contents

Advertisement

Hewlett-Packard Model 3478A
Digital Multimeter
Serial Number_____________
Ohms Test
Step#
Input to
3478A
1
Open
2
Short
3
Short
4
Short
5
Short
6
Short
7
Short
8
Short
9
30 ohm
(9)
10 ohm
10
300 ohm
(10)
100 ohm
11
3K ohm
(11)
1K ohm
12
30K ohm
(12)
10K ohm
13
300K ohm
(13)
100K ohm
14
3M ohm
(14)
1M ohm
15
30M ohm
(15)
10M ohm
PERFORMANCE TEST CARD
1 YEAR LIMITS
Set-Up and
Configuration
Press TEST/RESET
Ohms Function
00.0035
300 Range
000.004
3K Range
0.00002K
30K Range
00.0002K
300K Range
000.002K
3M Range
0.00002M
30M Range
00.0002M
30 Range
30.0104
10.0058
300 Range
300.018
100.009
3K Range
3.00013K
1.00006K
30K Range
30.0013K
10.0006K
300K Range
300.013K
100.006K
3M Range
3.00018M
1.00007M
30M Range
30.0110M
10.0038M
Test Performed By______________________
Date_____________
Reference Temperature__________________
High
Reading
Limit
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
_______
7
TM 11-6625-3071-14
Low
Test
Limit
Pass
00.0000
_____
000.000
_____
0.00000K
_____
00.0000K
_____
000.000K
_____
0.00000M
_____
00.0000M
_____
29.9896
_____
09.9942
_____
299.982
_____
099.991
_____
2.99987K
_____
0.99994K
_____
29.9987K
_____
09.9994K
_____
299.987K
_____
099.994K
_____
2.99982M
_____
0.99993M
_____
29.9890M
_____
09.9962M
_____
Test
Fail
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____
_____

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents