Glucose
Meter. Dropping or throwing the meter may cause
damage.
•
Don't expose the iHealth Smart Glucose Meter, test
strips, or control solution to extreme conditions such as high
humidity, heat, freezing cold, or dust.
•
Always wash your hands with soap and water, and rinse
and dry them completely before handling the iHealth Smart
Glucose Meter and test strips.
•
The iHealth system is not intended for use on neonates.
•
The iHealth system is not intended for use on artery
blood, serum, and plasma.
•
The iHealth system should only be used with iHealth
test strips.
•
The iHealth system can be used up to an altitude of
3276 meters(10744 feet).
•
The following substances at levels greater than normal
or therapeutic levels may cause significant interference (affect
the result by greater than 10%), resulting in an inaccurate
result: ascorbic acid, uric acid, acetaminophen, dopamine, L-
dopa, etc. These substances do not affect test results in normal
concen trations but may affect test results in high
concentrations. Do not use haemolysis samples, icterus
samples, or high lipemia samples.
•
Patients undergoing oxygen therapy may yield falsely
lower results.
LIMITATIONS OF USE
27