[1] When the meter is not connected to your iHealth Gluco-Smart
App on your iOS device:
Step 1: Turn on the control solution test (CTL) mode.
Insert the test strip into the meter's strip port to turn on the
meter. When the blood sample symbol appears on the meter
display, press the memory button to turn on the CTL mode.
When the CTL symbol appears, the meter is in control solution
test mode and will not save this test result in memory.
Note: Be sure to set the meter and/or app on the CTL mode before
performing a control solution test. The control solution test result
will not be saved in the meter.
Step 2: Press the memory button again to turn off the CTL mode
and switch back to the regular testing mode.
Step 3: Apply the control solution.
Shake the control solution vial before each use.
Squeeze a drop of control solution into the vial cap. For
better results, it is recommended that you use the second drop
of the control solution (discard the first drop).
Hold the meter and move the absorbent hole of the test
strip to catch the drop. Once the confirmation window fills
completely, the meter will start counting down. Remove the
22