control solution sample from the test strip when the
countdown begins.
Note: To avoid contaminating the entire vial of control solution, do
not directly apply control solution onto a strip.
Step 4: Read and compare the results.
After the meter counts down to "1," the control solution test
result will appear on the meter display.
The result of the control solution test should be within the range
printed on the test strip vial label. If the test result falls outside the
specified range, repeat the test, carefully following the steps
above.
[2] When the meter is connected to your iHealth Gluco-Smart App
on your iOS device:
Step 1: Launch the app.
Step 2: Tap the CTL button to switch to control (CTL) mode.
Step 3: Follow Steps 3-4 above in section [1].
Step 4: The control solution test result will appear on the app on
your iOS device.
Out-of-Range Results
Results falling outside the specified range may be caused by:
•
An error in the test
•
Expired or contaminated control solution
•
An expired or contaminated test strip
•
Meter malfunction
23